MTS-MONO-ME 1.120

MTS-MONO-ME 1.120

National Registry Number 89778-23

Portable test system

The MTS-MONO-ME 1.120 is a compact portable test system designed for accuracy testing/verification/calibration of smart energy meters and measuring transducers.

It combines three instruments in one: a reference meter, a current and voltage source and a volt-mA calibrator/error calculator.

Accuracy class: 0.05

Purpose

The MTS-MONO-ME 1.120 is intended for verification and calibration of:
single-phase smart energy meters with power quality measurement function (accuracy class 0.2S)
Test scheme
single-phase measuring transducers with unified DC current and voltage signals
Test scheme

Application

The MTS-MONO-ME 1.120 can be used both in the laboratory and field conditions.

With the MTS-Service software, testing is performed automatically including generation of test reports (complete with environmental data according to ISO/IEC 17025-2009 as well).  

Basic technical data

Parameter

Range

Measurement error

Voltage

Ranges: Unom = 60, 120, 240 V

6 ... 288 V

% of reference

±0.02 % + ∆

Current

Ranges: Inom = 0.05; 0.1; 0.25; 0.5; 1; 2,5; 5; 10; 25; 50, 100 А 

1 mA ... 120 A

% of reference

±0.02 % + ∆

Frequency

40 ... 70 Hz

Absolute

±0.002 Hz

Phase angle

0 ... 360°

Absolute

±0.01°

Power factor

0.1 ... 1.0

Absolute

±0.005

Active power

0.1Unom ... 1.2Unom

0.1Inom ... 1.2Inom

% of reference

±0.05 % + ∆

: additional error

Parameter

Output setting range

In increments of

Value

Voltage

Ranges: Unom = 60/220 V

Distortion

Max output power per phase

3 ... 300 V

0.01 V

 

 

≤1 %

30 VA

Current

Ranges: Inom = 1, 10, 100 A

Distortion

Max output power per phase

0,0025 mA ... 120 A

0,0001 mA

 

 

≤1 %

60 VA

Frequency

45 ... 70 Hz

0.01 Hz

 

Phase angle

–180° ... +180°

±0.01°

 

Voltage and current harmonic composition

Harmonics

Interharmonics

 

2 ... 50

0.5 ... 50.5

 

 

Parameter

Range

Measurement error

Input

 

% of reference

DC voltage

–15...+15 V

0...15 V

±0.03 %

DC current

–7.5...+7.5 mA

0 ... 30 mA

±0.05 %

Output

 

Absolute

DC voltage

–10.5...+10.5 V

±0.002

DC current

–24...+24 mA

±0.005

Parameter

Value

Pulse level

5 ... 15 V

Frequency (max)

36 kHz

Pulse duration

>14 µs

Constant

1 ... 999 999 999 pulse/(kW • h)

AC power consumption

400 VA, or less

Mains supply

195 ... 253 V
47 ... 63 Hz

Operating temperature range

+10  to +35 °C

Weight

16 kg,or less

Benefits

Compact size (3 instruments in one unit)

PC with MTS-Service software installed (serves for automatic control of test procedures and generation of test reports) is included in the standard supply

Meter connection time is minimized

Accessories

УПС
UPS - 1.120
Quick Meter Connection Device for one single-phase electricity meter with plug-in current and voltage modules
ups-1.120-3
UPS - 1.120.3
Quick Meter Connection Device for three single-phase electricity meters with plug-in current and voltage modules
Scanning Head
Scanning heads for testing induction meters or meters with optical output
1Ethernet-1RS232 Converter
 
converter-2ethernet
Converter 2Ethernet-4RS232 with Power Adapter
An accessory for connecting up to 4 RS-232 devices to a PC via its Ethernet ports
AO-USB
AO-USB
Optical probe
TCM-02C
Time correction module
For calibration of meters with current clamps up to 3000 A (10 A input)
Thermo-Hygrometer
 
 

Documentation

 

Software