Accuracy class: 0.005
EDN-500 is a reference voltage divider which produces an output voltage by dividing the AC voltage on its input by a selected division factor.
![]()  | 
Low power electronic voltage transformers manufactured in compliance with IEC 60044-7-2010 | 
![]()  | 
Energy meters | 
![]()  | 
Power and voltage converters | 
| Examples of tested transformers | 
The divider is a high-precision component of a test system typically installed in an industrial or metrological lab and used for testing, calibration, or verification of measuring instruments.
The divider can be used as part of:
![]()  | 
PCS-ME 3.0 (secondary standard of AC power) | 
![]()  | 
TTS-ME 61850 (transformer test system for calibration of digital transformers) | 
![]()  | 
Transformer test systems for calibration of low power electronic transformers manufactured according to IEC 60044-7-2010 | 
| 
 Voltage measurement range  | 
 Measurement error in the frequency range   | 
 Measurement error in the frequency range   | 
||
| 
 Ratio  | 
 Angle  | 
 Ratio  | 
 Angle  | 
|
| 
 16 ... 480 V  | 
 0.005 %  | 
 0.2 min  | 
 0.02 %  | 
 0.2·(f/50) min  | 
| 
 Rated input voltage  | 
 Rated output voltage of   | 
 Measurement error  | 
||
| 
 Ratio  | 
 Phase  | 
|||
| 
 60 V  | 
 30 mV ... 8 V  | 
 0.01 %  | 
 0.5 min  | 
|
*A number of output voltages can be pre-selected by the customer from a range of 1.625 ... 6.5 V according to IEC 60044-7–2010 (1.625 – 2 – 3.25 – 4 – 6.5 V).