The Test System is designed for accuracy testing (calibration, verification) of energy meters and instruments measuring electrical quantities.
Accuracy classes: 0.02, 0.05
Single- and three-phase active and reactive energy meters of accuracy class 0.05 or less accurate with input ratings up to 120A and from 100 to 400 V | |
LPIT-compatible electronic meters of accuracy class 0.5S (or less accurate) | |
Single- and three-phase meters of accuracy class 0.05 with operating frequency range up to 450 Hz used to measure active, reactive, and apparent power/energy | |
Voltmeters (up to 1000 V), ammeters and measuring converters (up to 450 Hz) with unified outputs | |
Measuring instruments equipped with AC current probes up to 3000 A | |
Instruments measuring power quality as applied to Russian Standard GOST R 8.655–2009 | |
Instruments measuring DC current, voltage, and power (accuracy class 0.05 or less accurate, 100 A, 600 V). |
With use of the State Calibration Scheme, depending on the modification, the Test System can be certified as a 1st grade or 2nd grade reference standard for the instruments measuring:
Electrical quantities in the frequency range from 1 to 2500 Hz as stated by Rossstandart, Order No 1436 of 23 July 2021 | |
AC current in the ranges from 1Е-8 to 100 A and from 1Е-1 to 1Е6 Hz as stated by Rosstandart, Order No 668 of 17 march 2022 | |
AC voltage in the ranges up to 1000 V and from 1Е-1 to 2Е9 Hz as stated by Rosstandart, Order No 1705 of 18 August 2023 | |
DC current in the range from 1Е-16 to 100 A as stated by Rosstandart, Order No 2091 of 1 October 2018. |
The Test System can be applied to calibration of smart meters (additional parameters to be measured are selected by the customer)
Energomonitor-3.1KM (52854-13) | |
Energomonitor-3.1KM-E (86936-22) | |
Energomonitor-61850 (73445-18) |
The portable version of the Test System is intended for on-site testing (verification, calibration) of LPIT-compatible electronic meters of accuracy class 0.5S with integrated power quality measurement function (class S).
Error calculators VP-3.1 and scanning heads of SH series (if included in the scope of supply) provide for testing meters in the automatic mode.
The multifunctional meter suspension rack STB provides quick connection of tested meters to the test system.
An example of the Test System equipped with the meter suspension rack STB-10 (for testing 10 3-phase meters at a time):
To provide galvanic insulation when testing 5 or more meters with closed U-I links (including directly connected meters) the Test System can be equipped with isolating three-phase current transformers TTR-3.100.
To provide galvanic insulation when testing 3 or more single-phase meters of accuracy class 1.0 with closed U-I links the Test Bench can be equipped with isolating single-phase voltage transformers TNI-1.
To test measuring instruments equipped with AC current probes (split-core CTs) up to 3000 A, the Test System is supplied with calibrated coils of KT series (used to multiply test current).
To test measuring instruments with unified DC voltage or current outputs, the Test System is supplied with volt/mA calibrators Calmar-S (69182-17).
“ME-Service”“ME-Service” is a HW/SW set expanding the functionality of the Test System by adding the function of generating test reports according to ISO/IEC 17025.
Closest analogue:
- Test systems for testing electricity meters produced by MTE (EMH Energie-Messtechnik GmbH, Germany)
- Test systems for testing electricity meters MTS by ZERA (ZERA GmbH, Германия)
- Electricity meter test system ELMA-8315 (Applied Precision Ltd., Slovakia)
Mains supply
207 ... 253 V,AC power consumption
6000 VA, or lessOperating temperaturerange
From +18 to +28 °С (stationary)
Parameters of the Source |
Range |
Setting increment |
Frequency of first harmonic (f1) |
40 to 450 Hz |
0.01 |
RMS of 1st harmonic of phase voltage (U1) |
From 0.0001 to 12 V |
0.0001 |
RMS of 1st harmonic of phase voltage (U1) generated using voltage amplifier VA-3.1 |
From 6 to 576 V |
0.001 |
RMS of voltage produced by 1st harmonic of output current signal Ui1 |
From 0.0001 to 6 V |
0.00001 |
RMS of 1st harmonic of current (I1) generated using current amplifier CA-3.1 |
From 0.002 to 120 A |
0.0001 |
Harmonics of frequency k·f1 (k = 2 …19) in steps of 1.0 |
0 to 100% |
0.01 |
Harmonics of frequency k·f1 (k = 20 … 50) in steps of 1.0 |
0 to 50% |
|
Harmonics of frequency k·f1 (k = 0.5 … 50.50) in steps of 1.0 |
0 to 15% |
|
Phase shift between 1st harmonics of voltage in different phases 1st harmonic of voltage and 1st harmonic of current in the same phase 1st and nth harmonics of voltage in the same phase 1st and nth harmonics of current in the same phase |
From –179.99 to +180.00 ° |
0.01 |
Duration of voltage dip or swell (t) |
From 0.0001 to 12 s |
0.0001 |
RMS value of voltage dip (Umin) in % от U1 |
0 to 100% |
0.01 |
RMS value of voltage swell (Umax) in % от U1 |
100 to 200% |
0.01 |
Energoforma
Software for automatic control of test (verification or calibration) procedures when testing reference instruments (like Energomonitor) or similar measuring instruments
Enform-MTS
Software for automatic control of test (verification) procedures when testing electricity meters or various measuring converters with unified outputs using volt/mA calibrator/error calculator Calmar-S
EmCounter
Software intended for managing and documenting test results and for maintaining databases of tested devices
EfCalibr
Software used to adjust generator Energoforma-3.1E
Capable of testing smart meters |
|
Multi-position: up to 30 electricity meters can be tested at a time |
|
Both manual and automatic test procedures are supported |
|
With the software, it is possible to set test signals in the generator, read records from the internal memory, and configure automatic generation of test reports |